Guide System 24 APPLICATION Form drill SCREW-LINE, Cortical bone (optional) If during implant bed preparation it is shown that mainly cortical bone (bone qualities 1* and 2*) is present, the apical portion of the implant bed can be widened using the form drill SCREW-LINE, cortical bone (see “Pro- duct overview”). This has the effect of reducing the insertion torque of the implant. Prior to implant insertion the implant bed has to be rinsed with sterile physiological saline solution to remove titanium chips (generated through contact of the drill cutting edge with the guiding sleeve). *see [A] in section «Further documentation», page 32