Work Instruction GUIDE-SYSTEM

Guide System 26 The Guide System driver, manual/wrench or Guide System driver with ISO shaft is inserted up to the stop on the insertion post. After this, the lo- cked torque wrench or angled hand piece with torque limiter is placed on the driver. The implant is then screwed into final position in the bone. IMPORTANT NOTE The implant has reached the planned vertical end position when the shoulder of the insertion post rests on the top of the guide sleeve. After reaching the final position, the implant may not be rotated further in the template as this can lead to loss of primary stability. After the vertical end position has been reached, a longitudinal mark on the insertion tool (position corresponds to the position of the grooves in the internal configuration of the implant) should be oriented in vestibu- larly or - if the orientation is defined by the preoperatively fabricated in- terim prosthesis, in the direction of the mark on the top of the sleeve. If this is not the case, then fine adjustment is required. Fine adjustment of the implant position The implant position can only be finely adjusted when the temp- lated has been removed . For this purpose, the insertion post and the template must be removed. After that, reinsert and tighten the insertion post, attach the insertion tool incl. torque wrench and correct the groove position. Final position: Resting of the insertion tool on the upper side of the guide sleeve. Driver, manual/wrench Driver, ISO shaft Locked torque wrench Final position: Matching markings of insertion tool and guide sleeve APPLICATION