30 HYBRID RESTORATIONS WITH THE CAMLOG ® IMPLANT SYSTEM INSERTING THE BAR CONSTRUCTION The CAMLOG ® bar abutments are transferred from the working cast to the CAMLOG ® implants and screwed in definitely with the prescribed torque. The finished bar construction is transferred to the CAMLOG ® bar abut- ments and fixed definitely with 15 Ncmwith new unused prosthetic screws, using a screwdriver (hex). The newly created full denture is then inserted and checked for proper fit. SLEEVE FOR TITANIUM BONDING BASE, BURN- OUT (PASSIVE FIT) Cast bar version with prefabricated bar sleeve made of burn-out plastic (POM) for full casting technique and titanium bonding base as a retaining element for the implant. The Passive Fit System makes it possible to fabri- cate cast bars absolutely tension-free. For bar fabrication, the bar sleeve is placed over the titanium bonding base. After completing the bar, it is bonded to the implants with the titanium bonding bases. The plastic sleeve of the bar base can be shortened occlusally to the height of the prosthetic screw. The overall height of the plastic sleeve is 14 mm. EXAMPLE: WORKING CAST WITH CAMLOG ® LAB ANALOGS (TITANIUM ALLOY). WAX-UP The bar wax-up is created based on the planning on the burn-out bar base directly. The wax thickness over the plastic coping should be at least 0.3 mm. Do not cover the delicate edge of the base with wax. Prefabricated bar components made of wax / plastic can also be used to fabricate a pre- milled bar construction. Embedding, casting and devestment happen as described on page 28–29 "BASE FOR BAR ABUTMENT, BURN-OUT". Sleeve for titanium bonding base, burn-out, bondable (passive fit) BAR RESTORATIONS