CAMLOG Impression-Taking

Impression-taking, Bite Registration, and Temporary Restoration on CAMLOG ® Implants 8 | 9 OPEN IMPRESSION-TAKING METHOD The impression posts, open tray, are color-coded, equipped with an internal fixing screw, and can be used with the platform switching method. For the open impression-taking method, an individually fabricated impres- sion tray to be perforated on the extension of the implant axis to allow release of the fixing screw is required. The fixing screw is secured in the impression post with an O-ring and must be hand-tightened using the screwdriver only, both in the implant and in the lab analog. Impression post 10.0 mm O-ring Lab analog Impression posts PS, open tray, for platform switching O-ring NOTE Before removing the impression, loosen the screw and pull it back until the stop (O-ring) is felt. Otherwise, removal of the impression will be impossible due to axial divergence of the implants, or the impression itself will be deformed through excessive compression. 3.0 mm Impression posts, open tray The fixing screw is equipped with a break-off point. If space limitations are encountered, it can be shortened by 3.0 mm by breaking it off with a screw- driver, hex. Caution! Shorten extra-orally, only.

RkJQdWJsaXNoZXIy MTE0MzMw