CAMLOG SCREW-LINE Surgical procedure

21 CAMLOG ® SCREW-LINE IMPLANT PRE-DRILLING AND CONTROL AXIS ALIGNMENT A tapered pre-drill SCREW-LINE with a coronal diameter of 2.8 mm and apical diameter of 1.7 mm is available for the SCREW-LINE configuration. Recommended drilling speed: 600 rpm The depth marks on the drill match the implant lengths 7*, 9, 11 and 13 mm. The maximum drilling depth is 16 mm. For safety reasons, a depth stop matching the proposed implant length should be used. Further drilling is performed with the form drills. The axis alignment is controlled with the paralleling pin. 2.8 mm 2.8 mm 1.7 mm 1.7 mm Pre-drill SCREW-LINE Ø 1.7–2.8 mm, 600 rpm Pre-drilling Control of axis alignment Paralleling pin SCREW-LINE 2.0 mm * The lowest marking corresponds to a drilling depth of 7 mm and is of no significance for the CAMLOG ® SCREW-LINE implants.