CAMLOG SCREW-LINE Surgical procedure
22 CAMLOG ® SCREW-LINE IMPLANT FORM DRILLING Diameter- and length-specific form drills are available for each implant size. The form drills are color-coded and laser-marked. The form drills included in the surgery sets are supplied with a color-coded, removable depth stop. This must only be used with form drills SCREW-LINE. Depending on the specified drilling depth (implant length), the hole diame- ter is expanded progressively with the series of form drills until the inten- ded implant diameter is achieved. The small graduations in diameter achieve a gentle preparation of the bone. Recommended drilling speeds: Ø 3.3 mm 550 U/min Ø 3.8 mm 500 U/min Ø 4.3 mm 400 U/min Ø 5.0 mm 350 U/min Ø 6.0 mm 300 U/min Form drill SCREW-LINE Drill sequence in ascending order to the drill hole expansion up to the defined implant diameter. Ø 3.3 mm Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Ø 6.0 mm 9 mm 11 mm 13 mm 16 mm Length SURGICAL PROCEDURE