38 CAMLOG ® SCREW-LINE IMPLANT SURGICAL PROCEDURE TRANSGINGIVAL HEALING CAMLOG ® HEALING CAPS Use of the CAMLOG ® Healing caps supports the development of the periimplant soft tissue. CAMLOG ® Healing caps are available in three diffe- rent geometries: • cylindric • wide body • bottleneck The healing caps are color-coded to match the respective implant diame- ter. CAMLOG ® Healing caps are inserted into the CAMLOG ® SCREW-LINE im- plant and tightened manually using a screwdriver, hex. The healing cap rests on the machined implant shoulder and does fully cover it. CAMLOG ® HEALING CAPS PS FOR PLATFORM SWITCHING The CAMLOG ® Healing caps PS (cylindrical, wide body, bottleneck) are ta- pered in diameter on the apical end and enable the adaptation of soft tissue over the implant shoulder. IMPORTANT NOTE! If CAMLOG ® Healing caps PS are used for healing, the later prosthetic restoration, including the CAMLOG ® Impression taking, must use CAMLOG ® Prosthetic components PS for platform switching to prevent tissue injury! Connection CAMLOG® SCREW-LINE implant – CAMLOG® Healing cap Connection CAMLOG® SCREW-LINE implant – CAMLOG® Healing cap PS