Work Instruction COMFOUR®
14 OCCLUSALLY SCREW-RETAINED PROSTHETICS ON CAMLOG ® AND CONELOG ® BAR ABUTMENTS For impression taking, the cap is inserted into the driver for impression caps for bar abutments. The cap is retained through friction. PP-Ø: Prosthetic platform diameter *(available from Q2/2018) Driver for impression caps for bar abutments Then the impression cap is screwed hand-tight to the finally fixated bar abutment in the implant and the driver is withdrawn again. Impression tak- ing is performed directly over the prosthetic platform of the CAMLOG ® and CONELOG ® Bar abutments. Ø 3.3/3.8/4.3 mm Ø 5.0 mm A closed tray is suitable for impression taking. The impression is then taken with silicone or polyether impression material. After removing the impression, the impression cap remains on the bar abut- ment. ART. NO. ARTICLE IMPLANT Ø IN MM PP Ø IN MM HEIGHT IN MM MATERIAL J2129.4300 J2129.6000 J2129.4310* J2129.6010* Impression cap short for bar abutment, closed tray (bridge/ bar), partial light blue anodized, sterile Impression cap long for bar abutment, closed tray (bridge/ bar), partial light blue anodized, sterile 4.3 6.0 4.3 6.0 3.3/3.8/4.3 5.0 3.3/3.8/4.3 5.0 6.5 7.0 11.0 11.0 Titanium alloy Titanium alloy