Work Instruction COMFOUR®

27 OCCLUSALLY SCREW-RETAINED PROSTHETICS ON CAMLOG ® AND CONELOG ® BAR ABUTMENTS TEMPORARY OR FINAL RESTORATION WITH TITANIUM CAPS FOR BAR ABUTMENTS, FOR CROWN AND BRIDGE Titanium caps for bar abutments for crown restorations are fitted with an internal hex as antirotational mechanism, and without internal hex for bridge restorations and are made of titanium alloy (Ti6Al4V ELI). The caps are available for implant diameters 3.3/3.8/4.3 mm and 5.0 mm, can be shortened individually and are packed sterile with a light blue anodized prosthetic screw. ART. NO. ARTICLE IMPLANT Ø IN MM MATERIAL J2259.4301 J2259.6001 J2259.4302 J2259.6002 Titanium cap for bar abutment, for crown, incl. prosthetic screw light blue anodized, sterile Titanium cap for bar abutment, for bridge, incl. prosthetic screw light blue anodized, sterile 3.3/3.8/4.3 5.0 3.3/3.8/4.3 5.0 Titanium alloy Titanium alloy Can be shortened by 8.5 mm Height 14.0 mm IMPORTANT NOTES • Shorten extra-orally only. • During intraoral use, products must be secured in general against aspiration and swallowing. A temporary restoration with titanium caps can be fabricated optionally directly on the definitively in the implants inserted CAMLOG ® and CONELOG ® Bar abutments or on the bar lab analogs on the working model. The caps can be shortened occlusally by 8.5 mm up to and including the fourth chamfer.

RkJQdWJsaXNoZXIy MTE0MzMw