CONELOG PROGRESSIVE-LINE Surgical procedures

26 CONELOG ® PROGRESSIVE-LINE IMPLANT B) Transparent blister with Tyvek ® foil and primary label: The blister with the Tyvek ® foil represents the primary packaging, the contents of which are sterile - implant holder with implant and cover screw. Furthermore, the blister includes four self-adhesive patient labels. These can, for example, be used for the patient records, the implant pass, the letter of referral. For faster orientation, the diameter information is also highlighted in color here. C) Implant holder with implant and cover screw: The implant holder securely fixates the implant and the cover screw in the packaging. Both the implant and the cover screw can be released and removed via a simple click mechanism with the implant holder. In addition, the implant can be clearly identified in the implant holder after removal from the primary packaging: a) The implant diameter can be identified via the color-coding of the insertion post and the cover screw. b) A scale on the bottom side of the implant holder allows reading the length of the implant: the position of the titanium retaining plate on the scale gives the implant length 7, 9, 11, 13 and 16 mm. D) Mounted insertion posts: The implants are secured in the implant holder with a color-coded insertion post corresponding to the diameter. The insertion posts are mounted in the implant and can be pulled off easily from the implant after implantation without requiring further tools. Ø 3.3 mm Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Scale to determine the length of the implant SURGICAL PROCEDURE