CONELOG PROGRESSIVE-LINE Surgical procedures

31 CONELOG ® PROGRESSIVE-LINE IMPLANT PICKING UP THE INSERTION POST WITH THE ANGLED HANDPIECE Optionally, the insertion post can also be picked up directly with the ISO shaft handpiece insertion tool on an angled hand piece. The front part of the implant holder is held and then the insertion post is picked up with the handpiece insertion tool by applying pressure . During the pick-up process, observe the correct alignment of the three groove markings on the head of the insertion post and the insertion tool. Only after inserting the insertion tool on the insertion post, press the implant holder together at the rear section (see arrows in the illustration) to release the lock on the implant holder and thus the implant. Lift out the insertion post upwards in a straight line (do not kink). Observe the correct alignment during the pick-up process! NOTE It should be noted that picking up the insertion post with the insertion tool is done by applying pressure. This ensures secure retention of the insertion post in the insertion tool.