CONELOG SCREW-LINE surgical procedures

27 CONELOG ® SCREW-LINE IMPLANT B) Transparent blister with Tyvek ® foil and primary label: The blister with the Tyvek ® foil represents the primary packaging, the contents of which are sterile - implant holder with implant and cover crew. Furthermore, the blister includes four self-adhesive patient la- bels. These can, for example, be used for the patient records, the im- plant pass, and the letter of referral. C) Implant holder with implant and cover screw: The implant retention system securely fixates the implant and the cover screw in the packaging. Both the implant and the cover screw can be re- leased and removed via a simple click mechanism with the implant hol- der. In addition, the implant can be clearly identified in the implant hol- der after removal from the primary packaging: a) One side of the implant holder contains information which identifies the type of implant. b) The implant diameter can be identified via the color-coding of the in- sertion post and the cover screw. c) A scale on the bottom side of the implant holder allows reading the length of the implant: the position of the titanium retaining plate on the scale gives the implant length 7, 9, 11, 13 and 16 mm. D) Mounted insertion posts: The implants are secured in the implant holder with a color-coded inser- tion post corresponding to the diameter. The insertion posts are mounted in the implant and can be pulled off easily from the implant after implantation without requiring further tools. Ø 3.3 mm Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Information on the type of implant Scale to determine the length of the implant

RkJQdWJsaXNoZXIy MTE0MzMw