Implant Placement and Immediate Final Rehabilitation

The Screw Line implants for the Camlog Guide sys- tem come with a special implant mount that match- es the sleeve diameter. The different diameters are colour coded (Fig. 77). The implant mount also fea- tures a depth stop at its upper edge, which will come to rest against the edge of the sleeve once the planned insertion depth is reached. An implant driv- er in a torque ratchet was used to tighten the implant (Fig. 78). The primary stability of the implant was checked against the force required for insertion. When all implants had been placed and the retaining screws released, each implant mount and the surgical stent were removed (Figs. 79 to 81). The insides of the implants were cleaned and covered with an anti- bacterial gel. After screwing the bar abutments refined by the laboratory to the implants, the wounds were sutured closed. This completed the implant procedure (Fig. 82). To verify the fit of the EDI Case Studies 18 Fig. 77 Camlog Screw Line implant with a Camlog Guide implant mount placed into an implant driver and a torque ratchet. Fig. 80 Removing the released implant mount. Fig. 79 Overview of implants placed. Fig. 82 Bar abutments screwed in place. Fig. 78 Fully inserted implant. Fig. 81 Intraoral overview after implant placement and template removal.