CAMLOG Hybrid Restorations
46 HYBRID RESTORATIONS WITH THE CAMLOG ® IMPLANT SYSTEM INSERTION OF THE CAMLOG ® BALL ABUTMENT AND THE PROSTHESIS After removing the CAMLOG ® healing cap, the CAMLOG ® ball abutments are transferred from the working cast to the previously cleaned CAMLOG ® implant and the driver for ball abutment and the torque wrench are used to tighten the abutments in the implants based on the specified tightening torque. CAMLOG ® abutments must be retightened with the same torque after about five minutes to reach the maximum screw tension. This prevents screws from loosening to the extent possible. The screwdriver/activator for ball abutment matrix is used to set the re- quired retention force of the lamella retention inserts and the prosthesis is inserted in the patient's mouth. IMPORTANT NOTE Do not place the stabilizing ring in the mouth. The clinical insertion is completed with a check of the occlusion and articu- lation. TIGHTENING TORQUE FOR CAMLOG ® BALL ABUTMENTS Ø 3.3 mm 20 Ncm Ø 3.8/4.3/5.0 mm 30 Ncm BALL ABUTMENT ANCHORING SYSTEM