CAMLOG SCREW-LINE Surgical procedure
24 CAMLOG ® SCREW-LINE IMPLANT CHECKING THE IMPLANT BED Probing the implant bed hole for fenestration is recommended. Results of probing tests for the absence of soft tissue in the implant bed hole must be documented in the patient file. Cortical bone drilling Ø 4.3 mm for implant length 13 mm Checking the implant bed CORTICAL BONE DRILLING If the bone quality is class 1* the cortical bone drill enables reduced-tor- queimplant insertion through controlled circular expansion of the implant bed. The flattened drill tip serves as the depth stop. A color-coded laser-marked cortical bone drill is available for each implant diameter. Form drill SCREW-LINE Cortical bone Ø 3.3 mm Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Ø 6.0 mm 550 500 400 350 300 Recommended drilling speed in rpm: SURGICAL PROCEDURE * Bone quality as documented in Lekholm U, Zarb GA. Patient selection and preparation. In: Branemark PI, Zarb GA, Albrektsson T, editors. Tissue-integrated prostheses-Osseointegration in Clinical Dentistry. Chicago: Quintessence Publishing Co; 1985. p. 199–209