CAMLOG SCREW-LINE Surgical procedure

25 CAMLOG ® SCREW-LINE IMPLANT Tapping in the upper region of the implant bed TAPPING All CAMLOG ® SCREW-LINE implants come with a self-tapping thread. Use of a tap is recommended for bone quality categories 1* and 2*. The maximum speed must not exceed 15 rpm when performing power- assisted tapping. We recommend manual tapping. Tap SCREW-LINE, with hexagon, max. 15 rpm Ø 3.3 mm Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Ø 6.0 mm Locked torque wrench Manual tapping is performed with tap adapters for the tap SCREW- LINE and the locked torque wrench. Make sure to pay attention to the axial di- rection of the implant bed when inserting and removing the tap. The limit for insertion of the tap is the upper edge of the cutting blade. Tap adapter, short and long, for tap SCREW-LINE * Bone quality as documented in Lekholm U, Zarb GA. Patient selection and preparation. In: Branemark PI, Zarb GA, Albrektsson T, editors. Tissue-integrated prostheses-Osseointegra- tion in Clinical Dentistry. Chicago: Quintessence Publishing Co; 1985. p. 199–209