33 CAMLOG ® SCREW-LINE IMPLANT IMPLANT INSERTION AND POSITIONING Using the insertion tool, the implant is inserted into the implant bed and carefully screwed in clockwise either manually or with the angled hand piece (maximum speed may not exceed 15 rpm). Pay attention to the axial alignment of the implant bed. In the case of manual insertion, once the implant has been inserted into the implant bed, the implant can be screwed into its final position with the wrench. If the thread was tapped in advance, the positions of the threaded ends in the cortical bone and on the implant must match. It is recommended to first rotate the insertion tool with the implant care- fully to the left manually, until the thread socket can be felt. Then the im- plant is screwed in clockwise manually with the insertion tool. When reaching the planned insertion depth (see section on form drilling), one of the 3 grooves should face in a vestibular direction. If it was decided to set preparation depths for the implants individually by removing the depth stop during form drilling, this must be kept in mind when inserting the implant. It is possible to individually position implants vertically to match the drilling depth. Insertion of implant with a manual insertion tool Screw insertion of implant with manual insertion tool and wrench (max. 15 rpm) Manually screwed in implant Insertion of implant with a machine insertion tool Screw insertion of implant with a machine insertion tool and angled hand piece (max. 15 rpm) Machine screwed in implant