CONELOG PROGRESSIVE-LINE Surgical procedures

24 CONELOG ® PROGRESSIVE-LINE IMPLANT Tapping in the upper region of the implant bed ALTERNATIVE Taps can also be used as an alternative to the dense bone drill. These are inserted into the cavity at different depths depending on the implant length. The marks on the taps represent the length-specific insertion depths (not proportional to the implant length) for 7 mm, 9 mm, 11 mm, 13 mm and 16 mm implants. The maximum speed of 15 rpm must not be exceeded with automated tapping. Manual tapping is recommended. Tap PROGRESSIVE-LINE, max. 15 rpm Ø 3.3 mm Ø 3.8 mm Ø 4.3 mm Ø 5.0 mm Locked torque wrench The adapter ISO shaft and the locked torque wrench are used to manually tap the thread. Make sure to pay attention to the axial direction of the implant bed when inserting and removing the tap. Adapter ISO shaft 13 mm 11 mm 7 mm 0 mm 16 mm 9 mm SURGICAL PROCEDURE