CONELOG PROGRESSIVE-LINE Surgical procedures
27 CONELOG ® PROGRESSIVE-LINE IMPLANT E) Insertion tools The implant can be picked up directly with the insertion tool via the mounted insertion post and removed from the implant holder. One of the five illustrated insertion tools can be used for this purpose. The insertion posts and insertion tools are designed such, that they are also suitable for narrow gaps: none of the components required for inserting the implant have a diameter greater than that of the implant itself. Furthermore, the long insertion tools also allow the placement of implants in narrow and deep anatomical situations. The figure illustrates the use of a handpiece insertion tool (with ISO shaft) with insertion post for the CONELOG ® Implant Ø 3.3 mm under tight interdental conditions. Insertion tool with ISO shaft The three manual insertion tools for use with the wrench (long, short, extra short). The two insertion tools with ISO shaft (long and short) for use with the angled hand piece. Insertion post Groove markings on the insertion tool and insertion post F) Insertion aid: If low primary stability is expected in a sinus lift procedure or in soft bone, then CAMLOG recommends the assembly of the insertion aid (see page 34) in place of the pre-mounted attached insertion post. The insertion aid is screw-retained as compared to the attached insertion post, and allows intraoperative corrections in positioning of the implant in all three spatial dimensions.
RkJQdWJsaXNoZXIy MTE0MzMw